info@tempcontrol.nl | +31 (0) 15 251 18 31

Stability Testing Climatic Chambers - Kambic Metrology

  • ST_klimaatkamer_1.jpg
  • ST_klimaatkamer_2.jpg
  • ST_klimaatkamer_3.jpg
  • ST_klimaatkamer_4.jpg
  • ST_klimaatkamer_1.jpg
  • ST_klimaatkamer_2.jpg
  • ST_klimaatkamer_3.jpg
  • ST_klimaatkamer_4.jpg

Stability Testing Climatic Chambers - Kambic Metrology

Perfect tool for stability testing in pharmaceutical Industry
• Following ICH guidelines
• Calibrated at all ICH points:

o 25 °C ± 2 °C / 60% Rh ± 5%
o 30 °C ± 2 °C / 65% Rh ± 5%
o 40 °C ± 2 °C / 75% Rh ± 5%
o 25 °C ± 2 °C / 40% Rh ± 5%
o 30 °C ± 2 °C / 35% Rh ± 5 %
o 40 °C ± 2 °C / 20% Rh ± 5%

• GMP Qualification package available
• Two different volumes available

Main features

  • Temperature range (°C): +10 ... +60
  • Temperature display resolution (°C): 0.1
  • Temperature set resolution (°C): 0.1
  • Temperature stability (°C): better than ±0.5°@ all ICH climatic points
  • Temperature uniformity (°C): better than ±1,5°C @ all ICH climatic points
  • Relative humidity range (%): 10...90
  • Relative humidity display resolution (%): 1
  • Relative humidity set resolution (%): 1
  • Relative humidity stability: better than ±3%@ all ICH climatic points
  • Temperature control: PID
  • Relative humidity control: Dew point humidity control
  • Access port (mm): Ø 40 standard,(Ø 50, Ø 90 as accessory)
  • Kambic Metrology