Stability Testing Climatic Chambers – Kambič Metrology
Perfect tool for stability testing in pharmaceutical industry
Features
- Following ICH guidelines
- Calibrated at all ICH points:
- 25 °C ± 2 °C / 60% Rh ± 5%
- 30 °C ± 2 °C / 65% Rh ± 5%
- 40 °C ± 2 °C / 75% Rh ± 5%
- 25 °C ± 2 °C / 40% Rh ± 5%
- 30 °C ± 2 °C / 35% Rh ± 5 %
- 40 °C ± 2 °C / 20% Rh ± 5%
- GMP Qualification package available
- Two different volumes available
Properties
- Temperature range (°C): +10 ... +60
- Temperature display resolution (°C): 0.1
- Temperature set resolution (°C): 0.1
- Temperature stability (°C): better than ±0.5°@ all ICH climatic points
- Temperature uniformity (°C): better than ±1,5°C @ all ICH climatic points
- Relative humidity range (%): 10...90
- Relative humidity display resolution (%): 1
- Relative humidity set resolution (%): 1
- Relative humidity stability: better than ±3%@ all ICH climatic points
- Temperature control: PID
- Relative humidity control: Dew point humidity control
- Access port (mm): Ø 40 standard,(Ø 50, Ø 90 as accessory)
Downloads